Force Calibration & Proof-Load Documentation

Agree the report scope before ordering: factory test records, accredited calibration and proof-load documentation answer different engineering questions.

Review the configuration and documents →
Force Calibration & Proof-Load Documentation engineering process schematic
ENGINEERING SCHEMATIC · NOT A MODEL DRAWING OR TEST RECORD

Choose the report that answers the acceptance question

DocumentUseful evidenceWhat it does not establish alone
Factory force / hysteresis recordIdentified sensor, direction, applied loads, readings, zero return, environmental conditions and reference equipment.Independent laboratory accreditation or an entire testing-machine qualification.
ISO/IEC 17025 accredited calibration reportIdentified issuing laboratory and current scope, method, range, uncertainty and traceable results.All methods/ranges or approvals outside that laboratory's scope.
Proof-load test reportDevice serial, actual test load, duration, method, inspection and acceptance result.Minimum breaking load, fatigue life or a blanket complete-lifting-system approval.

Specify tension and compression as separate calibration directions

For a bidirectional measurement chain, identify the required loads and acceptance criteria in each direction. Agree the transducer, readout, cable, gain and software scaling covered by the report. Changing those components can change the measurement chain.

ASTM E4 concerns testing-machine force calibration and verification. ISO 376 concerns force-proving instruments used to verify uniaxial testing machines. Request the applicable edition, scope and report from the responsible provider rather than inferring compliance from a brand label.

Keep WLL, proof load and breaking load in separate fields

For a requested 200% WLL proof test, define the actual model, method, load duration and acceptance criteria in the order. A test at twice WLL does not measure the minimum breaking load or establish a 5:1 / 10:1 design factor. Agree the applicable rigging standard edition and application requirements with the responsible qualified parties.

REPORT SCOPE / FIELD CHECKLIST

Know what a supplied report must contain

Preview the information to request before ordering. This is a blank field checklist, not an issued calibration certificate, proof-test result or laboratory accreditation.

Download open checklist PDF →

Report fields to request

No test values, certificate number, issuer seal or accreditation status is represented here.

Factory force record

  • Sensor model / serial and direction
  • Reference instrument and traceability
  • Actual applied loads and readings
  • Loading / unloading, zero return and repeatability
  • Environmental conditions, date and sign-off

Accredited calibration report

  • Issuing laboratory identity
  • Accreditation body, certificate and current scope
  • Force range, direction, method and uncertainty
  • Instrument identification and calibration results
  • Any limitations and authorised report signatory

Proof-load test record

  • Device model, serial and rated working load
  • Test method, applied load and duration
  • Reference loading equipment and traceability
  • Inspection / acceptance criteria and actual outcome
  • Breaking-load evidence separately identified

Verify the issuer before treating a report as accredited

Request the laboratory's identity and accreditation scope, then check the issuing accreditation body's current record. Keep the actual report issuer and signatory visible. Customer or order details may be redacted where permitted; the issuer is not relabelled as a TLC-owned laboratory.

Resolve the configuration before the purchase order

Does a factory force record establish laboratory accreditation?

No. A factory record describes the stated test. Accredited calibration requires an identified laboratory with a current accreditation scope covering the method, range and uncertainty needed.

Can a proof-load report replace a breaking-load report?

No. A proof test and minimum breaking-load evidence answer different questions. Specify the test scope, acceptance criteria and mechanical design evidence separately.

CONFIGURATION / INTERFACE / DELIVERY

Send an engineering RFQ with the details intact

Choose the force path, signal and matching accessories. Include the destination so the quotation can confirm the product, calibration, freight and requested delivery term together.

[email protected]

Open engineering notes are available in the downloads centre. Approved model STEP, IODD and real calibration reports are supplied against the selected configuration. Price, availability, lead time and warranty terms are confirmed in the formal quote.

1 / Contact
2 / Sensor geometry and configuration
6 / Hardware & calibration requests

Select a geometry to see matching assembly options.

Options are quotation requests. The final assembly rating and documentation scope require model-specific confirmation.

8 / Quantity & delivery country

TLC Tensile Metrology uses these details to respond to your enquiry. Privacy policy.

Review or select the complete RFQ text

The web form sends text. To include a drawing, datasheet or photograph, attach it in your email client and send to [email protected]. Selecting, copying or downloading a draft does not submit it. This form does not save a draft in browser storage.